Atomic Absorption Spectrophotometer, Auto Set Spectral Bandwidth

Model : SP-AA1800

Wavelength Range : 190-900nm
Wavelength Accuracy  : ≤0.15nm
Wavelength Receptivity : ±0.1nm
Size(W×D×H)(mm) : 1250x795x765(Main instrument)
The atomic absorption spectrophotometer speeds up analysis by eliminating the need to manually adjust the bandwidth.
Availability:
Quantity:

Atomic Absorption Spectrometry (AAS) is a simple and rapid technique for analyzing elements in solution.

The atomic spectrophotometer is commonly used in water, beverages, mining operations, clinical research, and pharmaceutical analysis.

What is an atomic absorption spectrophotometer?

An atomic spectrophotometer is an analytical instrument that uses Atomic Absorption Spectroscopy (AAS) to determine the concentration of metallic elements in a sample. The elements in a liquid or solid sample are detected by different wavelengths of radiation absorbed by different atoms.

Atomic Absorption Spectrophotometer Specifications:

ModelSP-AA1800
Light SourceSingle-element or multi-element hollow cathode lamp
Lamp HolderAutomatic switching of six-lamp platform and full-automatic collimation
Lamp CurrentPulse power supply
Optical SystemLarge area 1800 /mm reticle grating, fully enclosed optical system
Wavelength Range190-900nm, automatic peak finding, one-button optical optimization
Wavelength Accuracy≤0.15nm
Wavelength Repeatability±0.1nm
Spectral Bandwidth0.1/0.2/0.4/1.0/2.0nm Auto Set
Baseline DriftStatic ≤ 0.002 A/30 min, dynamic ≤ 0.005 A/30 min
Absorbance Range0-4A
DetectorImported photomultiplier tube
Size(W×D×H)(mm)1016x520x576
Shipping Size(W×D×H)(mm)1270x650x720
N.W./G.W. (kg)100/125
Flame System
Burner HeadAll titanium burner head, 50mm or 100mm universal burner head
Atomizing ChamberMacromolecule anti-explosion and anti-corrosion atomizing chamber
NebulizerHigh-efficiency glass nebulizer, can also be customized
Ignition ModeMicrocomputer control, automatic ignition
Gas ControlAutomatic gas control system
Characteristic Concentration0.015µg/mL/1%(Cu)
Detection Limit0.002µg/mL(Cu)
PrecisionRSD≤0.5%
SafetyMultiple protection measures such as gas leakage alarm, automatic protection
against tempering, and automatic power failure in case of abnormality
Graphite Furnace System
Heating ModeLongitudinal heating
Temperature Control ModeLongitudinal optical temperature control monitors the inner wall temperature of the graphite tube
Temperature RangeRT to 3000 c
Program Temperature ControlAutomatic temperature control up to 20 stages, enrichment and concentration in the furnace up to 20 times
Characteristic Quantity0.5×10-12g (CD)
Detection Limit0.4×10-12g (CD)
PrecisionRSD≤3%
Cooling WaterOptional cooling water circulation system
SafetyGraphite tube damage, water flow, air pressure alarm;
Water temperature overheat protection
Graphite Furnace Autosampler (optional)
Sample Tray130-position sample cup, 6-position reagent cup
Injection Volume0.01-100µl
Minimum Increment0.01µl
Injection Volume Repeatability0.01
Duplicate InjectionsUP to 99
Wash Container Volume500mL
Background Correction
Deuterium Lamp Background Correction1A background can be corrected
Self-priming Background Correction1A background can be corrected
Data Processing
Measurement MethodsFlame method, graphite furnace method, hydride generation-atomic absorption method
Calculation Methods of ConcentrationStandard curve method (cubic curve), automatic fitting, and standard addition method
Repeat Measurement Times1–99 times, calculate the average value, and give the standard deviation and relative standard deviation
Result PrintingParameter printing, data result printing, graphic printing, and word and excel documents can be exported

Atomic Absorption Spectrophotometer Features:

  1. SP-AA1800 atomic absorption spectrophotometer is widely used in scientific research, environmental protection, quality inspection, and other industries. The following are its features:

  2. High precision, fully automatic optical system, all lenses are coated with quartz coating to ensure analysis accuracy.

  3. Polymer material anti-corrosion atomization chamber to ensure sensitivity and stability.

  4. Optional 50mm and 100mm titanium burner greatly improves the flame efficiency and analysis accuracy.

  5. Fully automatic analysis can be completed with safe ignition, extinguishing and switching operations to improve the sensitivity of flame analysis.

  6. Graphite furnace temperature control can increase the highest temperature to 3000 ℃ / s. With automatic optimization, it can automatically ignite and adjust the flame height.

  7. SP-AA1800 atomic spectrophotometer test sensitivity reaches the advanced level in the industry, with sensitivity ≤ 0.015μg/mL/1%.


Previous: 
Next: 
Get Quotation/Become Distributor

About Company​​​​​​​

As a professional lab and medical equipment manufacturer, Scitek Global is certified by ISO 9001, ISO 13485, ISO 45001 and ISO 14001. Almost all our products are certified by ETL, CE and FDA .

Scitek Global looks forward to establishing cooperation with more distributors all over the world, and working together to create greater customer value.

Quick Links

Contact Us
Scitek Global Co.,Ltd
  No. 7000 Jingshi Road, Jinan City, Shandong Province, China
/  +86-531-82893797
  Inquiry: info@scitekglobal.com
Service: 
service@scitekglobal.com
Complaints: 

complaints@scitekglobal.com
 
Scitek Global Inc.
9999 Bellaire Blvd Ste 505, Houston, TX 77036, United States
  +1 (915) 4126735
info@scitekglobal.net
 
Scitek Ethiopia
 tsion@scitekglobal.com
Copyright © 2023 Scitek Global Co., Ltd. All rights reserved.  Sitemap  Privacy Policy