Horizontal Pulse Vacuum Autoclave

Model : ST-BVB150PV/ 250PV/300PV/360PV/450PV/600PV/800PV

Interior Design Pressure : -0.1~03MPa
Jacket Design Pressure : 0.3MPa
Design Temperature 150℃
Sterilization Temperature Control Precision : 0-2℃
Number of Positive Pressure Pulsations : 1~3
Air Filter : Filtration accuracy is <0.22 microns, bactericidal rate of 99.97%.
Horizontal pulse vacuum autoclave is designed to meet today’s changing lab sterilization needs.
Availability:
Quantity:

The ST-BVB series pulse vacuum autoclave employs saturated steam as its medium, employing pulse vacuum to thoroughly eradicate microorganisms and spores from the surfaces of articles. This method stands out as the most effective approach to physical sterilization, ensuring a highly reliable sterilization effect. The pulse vacuum exhaust successfully eliminates the impact of cold air on temperature, enabling sterilization through a combination of vacuum dehumidification and jacket drying processes.

Horizontal Pulse Vacuum Autoclave Specifications:

ModelST-BVB150PVST-BVB250PVST-BVB300PVST-BVB360PVST-BVB450PVST-BVB600PVST-BVB800PV
Capacity (L)150250300360450600800
Interior Design Pressure-0.1~03MPa
Jacket Design Pressure0.3MPa
Design Temperature150℃
Max Working Temperature139℃
Max Working Pressure0.25MPa
Vacuum Leakage Rate≤0.13KPa/min
Final Vacuum-96KPa
Opening Pressure of the Inner Chamber Safety Valve0.28MPa
Jacket Safety Valve Opening Pressure0.28MPa
Sterilization Temperature Control Precision0-2℃
Temperature Uniformity±0.5℃
Negative Pressure Pulsating Vacuum Negative Value and NumberAmplitude -80~0KPa, frequency 1~99
Number of Pulsations Across PressureAmplitude -80 ~80KPa, frequency 1
Number of Positive Pressure Pulsations1~3
Inner Tank Size (WxH)(mm)450x450550x550550x550610x610650x700650x700650x700
Depth75085010001000100013201750
Overall Size (LxWxH)(mm)

1050x1200

x1750

1150x1300x1800

1300x1300

x1800

1300x1350

x1900

1300x1360

x1950

1620x1360

x1950

2050x1360

x1950

Weight (kg)550850850950105011501350
Gas Consumption (Kg/c)12181822303645
Power Supply380V,50Hz380V,50Hz380V,50Hz380V,50Hz380V,50Hz380V,50Hz380V,50Hz
Consumption2KW+(12KW)2KW+(24KW)2KW+(24KW)2KW+(30KW)2KW+(30KW)2KW+(48KW)2KW+(48KW)

Horizontal Pulse Vacuum Autoclave Application:

This pulse vacuum autoclave is versatile and finds extensive application in pharmaceutical biological engineering, medical health, laboratory animal research, and other related fields. Specifically designed for items with stringent sterilization requirements, it is ideally suited for sterilizing utensils, aseptic clothing, rubber plugs, aluminum caps, raw materials, filters, media, and various other high-demand sterilizable items.


The utilization of saturated steam in conjunction with pulse vacuum not only enhances the efficiency of sterilization but also contributes to the overall reliability of the process. The pulsating vacuum exhaust ensures a thorough elimination of cold air, creating optimal conditions for the sterilization of diverse items. With its advanced technology and broad applicability, stands as a dependable solution for meeting the rigorous sterilization needs across a spectrum of industries.

Horizontal Pulse Vacuum Autoclave Features:

  1. Translation door, with anti-pinch hand function. Vacuum autoclave also adopts a double-door channel design, which is not only equipped with a powerful pressure safety interlock function but also a temperature safety interlock on the liquid process door to guarantee the safety of operation in all aspects.

  2. The high-efficiency design of the automatic sealing ring makes its sealing effect outstanding. The service life is up to 1500 times, providing users with long-lasting and stable performance.

  3. The main structure is european ring reinforcement jacket structure, programmable controller PLC, color touch screen.

  4. Provides a wealth of customized procedures, covering aseptic clothing, utensils, rubber, liquids, media, waste, sterilization, and many other areas, but also includes BD test, vacuum leakage test, and other professional test procedures to meet the needs of different areas.

  5. Pulse vacuum autoclave has a built-in micro-needle printer and a paperless recorder to make recording and tracking more convenient.

  6. Automatic drainage device, optional jacket automatic control drainage device, the inner chamber are installed automatic control drainage device, automatic detection jacket and inner chamber temperature pressure automatic drainage, to ensure the saturation of steam.

  7. Intelligent maintenance system: Optional remote monitoring and maintenance module; Water saving noise reduction system is optional

WHY SCITEK

  1. 1-2 Days Handling and 7-30 Days Shipping

  2. 1 Year Limited Warranty

  3. CE Standards


Previous: 
Next: 
Get Quotation/Become Distributor

About Company​​​​​​​

As a professional lab and medical equipment manufacturer, Scitek Global is certified by ISO 9001, ISO 13485, ISO 45001 and ISO 14001. Almost all our products are certified by ETL, CE and FDA .

Scitek Global looks forward to establishing cooperation with more distributors all over the world, and working together to create greater customer value.

Quick Links

Contact Us
Scitek Global Co.,Ltd
  No. 7000 Jingshi Road, Jinan City, Shandong Province, China
/  +86-531-82893797
  Inquiry: info@scitekglobal.com
Service: 
service@scitekglobal.com
Complaints: 

complaints@scitekglobal.com
 
Scitek Global Inc.
9999 Bellaire Blvd Ste 505, Houston, TX 77036, United States
  +1 (915) 4126735
info@scitekglobal.net
 
Scitek Ethiopia
 tsion@scitekglobal.com
Copyright © 2023 Scitek Global Co., Ltd. All rights reserved.  Sitemap  Privacy Policy